10 000 000 ₽
Automated setup for measuring thermoelectric properties in nanofilms STN 300.600.3Omega. The setup is designed to measure the Seebeck coefficient, electrical conductivity, and thermal conductivity coefficient in nanofilms. The thermal conductivity coefficient in nanofilms is measured using the 3-omega method.
Technical specifications:
Measurement temperature range: from room temperature to 300 °C.
Settable temperature gradient: from 3 °C, with a 1 °C step.
Sensitivity for the Seebeck coefficient: 1 µV/K.
Electrical conductivity measurement range: from 10E-9 to 1000 S/m; the range can be extended.
Thermal conductivity coefficient measurement range: from 10E-3 to 300 W/(m ·K).
The setup for measuring thermoelectric properties consists of the following main components:
a) personal computer;
b) digital oscilloscope;
c) control and measurement unit;
d) measurement vacuum chamber, also referred to as the measurement station;
e) vacuum pump.
The control and measurement unit of the setup includes:
— a module with an analog-to-digital converter (ADC) and a digital-to-analog converter (DAC);
— an electronic board with a galvanically isolated power supply for amplification of input and output analog signals;
— two PID controllers with RS-485 interface. The accuracy class of the PID controller is 0.5;
— a vacuum gauge.
The measurement vacuum chamber, or measurement station, structurally consists of a steel disk-shaped base, a glass bell jar, and two heating stages. The steel base has three vacuum feedthroughs. Each measurement stage is heated by radiation from a halogen lamp. The measurement vacuum chamber enables measurements from room temperature to 300 °C. The measurement vacuum chamber includes two platinum resistance temperature detectors for measuring the temperature of the stages. The vacuum pump ensures pressure in the measurement vacuum chamber of no more than 1 mmHg.
The software of the setup controls the control and measurement unit for performing measurements, including experimental data recording and graphical visualization.
Broadband EMI Shielding Performance in Optically Transparent Flexible In2O3/Ag/In2O3 Thin Film Structures. Materials 2025, 18(23), 5393.
https://doi.org/10.3390/ma18235393
The STN 300.600.3Omega setup for measuring thermoelectric properties is an in-house development of LLC Research and Production Company “Spectehnauka”.
Automated setup for measuring thermoelectric properties. Thermoelectric properties measurement system. Setup for Seebeck coefficient measurement. Nanofilm thermoelectric measurement. Thermoelectric characterization of thin films. STN 300.600.3Omega setup. In‑house development by Spectehnauka LLC. Seebeck coefficient measurement. Thermal conductivity coefficient measurement. Electrical conductivity measurement (nanofilms). 3‑omega method / 3ω method. Simultaneous measurement of Seebeck, electrical and thermal conductivity. Temperature range: room temperature to 300°C. Settable temperature gradient from 3°C, 1°C step. Seebeck coefficient sensitivity: 1 µV/K. Electrical conductivity range: 1E‑9 to 1000 S. Extendable electrical conductivity range. Thermal conductivity range: 1E‑3 to 300 W/(m·K). Pressure in vacuum chamber: ≤1 mmHg. Measurement vacuum chamber. Glass bell jar with steel base. Two heating stages (halogen lamp radiation). Platinum resistance temperature detectors (Pt RTDs). Vacuum pump. Control and measurement unit. ADC / DAC module. Galvanically isolated power supply for analog signals. Two PID controllers, RS‑485, accuracy class 0.5. Vacuum gauge. Digital oscilloscope. Personal computer for control. 3‑omega method for thermal conductivity. Measurement of thermal conductivity in nanofilms. Radiation heating of measurement stages. Vacuum chamber for thermoelectric measurements. Automated thermoelectric properties setup price. Buy STN 300.600.3Omega. Nanofilm Seebeck coefficient measurement system. Thermoelectric characterization for research. In‑house developed measurement setup for labs. Seebeck / electrical / thermal conductivity combo system. Nanofilm thermoelectric research. Material science laboratories. Thin‑film thermoelectric generator development. High‑temperature thermoelectric measurements (up to 300°C). R&D in physics of thin films. Control software for thermoelectric measurements. Graphical display of experimental data. Real‑time measurement logging. Accuracy class of controllers. Galvanic isolation in measurement paths. Custom development by Spectehnauka .
| Цена | от 2 200 000 руб. |
|---|